E-mail nk.trade@nihonika.co.jp
3-9, Tamatsukuri Motomachi, Tennoji-ku, Osaka 543-0014, Japan
■Each of the chambers can be individually controlled for temperature and illumination.
modelLP-30LED-8CTAR
modelLP-80LED-6CTAR
■Features
- It is compact, small-footprint, and energy-saving, and can be installed anywhere.
- Temperature can be controlled between +5 and +45°C, with changeover between constant operation and day/night switching operation. Each chamber can be individually controlled for temperature.
- Innovative air jacket system provides animal-friendly and stable, constant-temperature environment.
- Animals are housed in each chamber with less strass, and there is no cross-contamination among chambers.
- Each chamber is clean. HEPA filter is attached for inlet air and deodorant filter for exhaust air.
- Light intensity and lighting time can be individually controlled for each chamber, with changeover between constant operation and day/night switching operation.
- An inner door is installed to observe animals in each chamber.
- LP-30LED-8CTAR has 8 chambers with interior dimensions of W315 × 283 × H240 mm (small size).
- LP-80LED-6CTAR has 6 chambers with interior dimensions of W390 × D500 × H260 mm (large size).
■Specifications
Model LP-30LED-8CTAR LP-80LED-6CTAR Outer dimensions (mm) W1,255×D577×H1,900 W1,555×D825×H1,985 Inner dimensions (mm) W315×D283×H240×8 chambers W390×D500×H260×6 chambers Temperature +5 to 45℃ (+15 to 45℃ when light is on) Ventilation (air intake) 20L/min (8 chambers in total) with HEPA filter 60L/min (6 chambers in total) with HEPA filter Ventilation (air exhaust) 20L/min (8 chambers in total) with deodorant filter 60L/min (6 chambers in total) with deodorant filter Temperature control Changeover between constant operation and day/night switching operation. Illumination LED lamp x 8 chambers
Dimmable function (individually controlled), timer (individually controlled)
LED lamp x 6 chambers
Dimmable function (individually controlled), timer (individually controlled)
Power requirement Single phase 100V, 50/60Hz, 19A Single phase 100V, 50/60Hz, 18A
* These products should be used at an ambient temperature lower than 30°C.
* Temperature in a chamber may increase beyond the set value depending on radiation heat from lamps or heat generated
by animals.
* The lowest temperature may have to be set to 15°C or higher depending on lighting or temperature setting.