Nippon Medical & Chemical Instruments Co., Ltd. is a manufacturer specializing in environmental control, medical device, and scientific instruments.
Dimmable WLEDs (with a color rendering index of RA 90 or more) that generate low heat are installed in the ceiling as standard equipment.
Model | LH-61WLED | LH-61WLED-S | |
---|---|---|---|
Outer dimensions (㎜) | W434 × D446 × H982 | W603 × D531.5 × H982 | |
Inner dimensions (㎜) | W370 × D374 × H430 | ||
Control system | Changeover between constant operation and day/night switching operation | ||
Control methods | Three-position control method: Refrigerator and humidifier, ON/OFF control; heater, proportional control | ||
Temperature | +15℃ to 45℃ ± 2℃ ( +20℃ to 40℃ when all lights on) | ||
Illuminance1 | 0 to 8,100 lx | 0 to 14,600 lx | |
Photon flux density1 | 125 μmolm-2s-1 | 220 μmolm-2s-1 | |
Light source |
Illumination |
Ceiling: WLED, ≥ RA 90 |
Ceiling, left, right, rear sides: WLED, ≥ RA 90 Number of LED chips: 63(7 × 9) |
Number of lamps | Ceiling: 1 panel |
Ceiling, left, right, and rear: 4 panels |
|
Cooler | 100W | ||
Heater | 150W | ||
Operation current (maximum) | 2A | 2.5A | |
Power requirement | Sigle phase 100V 50/60Hz 15A | ||
Weight | 50㎏ | 60㎏ |
1 Illuminance and photon flux are measured on the center shelf inside the cabinet.
* Safety devices: Earth leakage circuit breaker, refrigerator protection device (overload relay [automatic recovery type]), and fuse
* The cooling device uses a CFC-free natural refrigerant.
* The temperature sensor: platinum resistance thermometer
* The power supply does not include the capacity of the outlet inside the cabinet. An optional outlet inside the chamber can be used, with a maximum capacity of up to 1A.
* The unit is designed to be used in a typical indoor environment with room temperatures from 15°C to 30°C. Depending on usage conditions, the specified performance may not be achieved.
* Drainage facilities may be required depending on the sample in the chamber.
* When lighting, the temperature should be set at 20°C or higher. Leave a minimum of 30 cm of space on both sides of the chamber; otherwise, it may lead to a deterioration in the cooling function.
* Accuracy: Temperatures are those when the lighting is off.
* If you operate this chamber at high temperatures for an extended period, please get in touch with our nearest sales office.