Products

With Dimmable LED

Plant Growth Chamber <High-light intensity model>

This is equipped with high-light intensity LED lamps.
The temperature and humidity are controlled by a 3-position control system, contributing to energy savings. Optimal program control can be achieved based on the specific requirements of the cultured plants. All types are capable of continuous dimming.

Features

  • An operation touch panel with time display is mounted on the upper part of the main body, which allows easy operation.
  • Model L(P)H-□PFQ (S/D/P) T-S allows both constant operation at any temperature/light/ (humidity) and day/night switching operation.
  • Model L(P)H-□PFQ (S/D/P) -SP(C) allows a program control.
  • Model L(P)H-□PFQ (S/D/P) -SPC has a CO2 control function.
  • Five-sided illumination. The light can be continuously dimmed form 20% to 100% or 0%.
  • Model LHs don’t need a drain hose or a tank because of a forced evaporation system inside the chamber.
  • Three-position controller is used for temperature and humidity control (proportional control for heating), which provides low running costs.
  • The inner door has energy-saving functions, being used as one door or two up-and-down doors as per usual.
  • Two-level wind speed control is installed to protect wind-sensitive plants..
  • A hole of 40 mm diameter for measurement is included as standard.
  • Humidity control system can be selected from “only humidification” or “only dehumidification” or “both humidification and dehumidification.”

Options

  • Table for a water tank (Model LPH) (Size:W300×D400×H900mm)
  • Evaporator for drained water (Model LPH)
  • Individual ceiling light control

LED Spectrum

 

Specifications

 

Model   White LED LH-411PFQST-S LPH-411PFQST-S LH-411PFQST-SP LPH-411PFQST-SP LPH-411PFQST-SPC
 Yellow LED LH-411PFQDT-S LPH-411PFQDT-S LH-411PFQDT-SP LPH-411PFQDT-SP LPH-411PFQDT-SPC
Pink LED LH-411PFQPT-S LPH-411PFQPT-S LH-411PFQPT-SP LPH-411PFQPT-SP LPH-411PFQPT-SPC
Outer dimensions (mm) W880 × D806 × H1,935
Inner dimensions (mm) W670 × D565 × H1,100
Control system

Changeover between constant operation
and day/night switching operation 

Real-time based control (24 hours)/Cumulated hours-based control(Up to 999 hours and 59 minutes/step)

24steps/5patterns with a pattern link function

Repeat 1 to 99 times or unlimited
 Control methods Three-position control method: Refrigerator and humidifier, ON/OFF control; heater, proportional control 
Temperature  +5℃ to 50℃ ± 1℃ ( +15℃ ~ when all lights on)
Humidity

50-90%RH ± 10%RH
(15℃ to 45℃)

50-90%RH ± 10%RH
(15℃ to 45℃)
CO2 concentration

Ambient to 3,000 ppm

Illuminance1 White: 0 to 58,500 lx,          Yellow: 0 to 76,000 lx,         Pink: 0 to 18,500 lx 
They can be continuously dimmed from 20% to 100% or 0% (all lights off)
Photon flux density1 White:0 to 900 μmolm-2s-1,      Yellow: 0 to 930 μmolm-2s-1,       Pink: 0 to 640 μmolm-2s-1

Light

source

Illumination

White: Straight tube sun-like LED lamp (SUNRAY LIGHT) 
 Yellow and Pink: Straight tube LED lamp (PLANTFLEC) for plant growth

Number of lamps  Left, right, rear, and door: 4 lamps (40W) x 4 sides, Celling: 1 lamp (600W) x 1 side
Refrigerator 300W 
Heater 500W
Humidifier 80W  80W 
Shelf 6 shelves (adjustable)  

Operation current
(maximum)

White: 17A 
Yellow and Pink: 16A

White: 20.5A 
Yellow and Pink: 19.5A

White: 17A 
Yellow and Pink: 16A

White: 20.5A 
Yellow and Pink: 19.5A

Power requirement

Single phase 100V
50/60Hz 20A E 

Single phase 100V
50/60Hz 30A E 

Single phase 100V
50/60Hz 20A E 

Single phase 100V
50/60Hz 30A E 

Weight (kg) About 295 About 305 About 295 About About 305

1 Illuminance and photon flux are measured on the center shelf inside the cabinet.

* Temperature sensor: platinum resistance temperature sensor, humidity sensor: capacitance measurement sensor, humidifier: ultrasonic humidifier.
* Power requirement does not include the power-outlet installed inside the chamber.

* These chambers are designed to operate in ambient temperatures up to 30℃. The specified performance may not be achieved depending on actual usage conditions.
* If lighting, temperature setting should be 15℃ or higher.
* The specified performance may not be achieved depending on the combination of temperature and humidity.
* Depending on the sample, drainage installation may be required.

RETURN TO TOP