Nippon Medical & Chemical Instruments Co., Ltd. is a manufacturer specializing in environmental control, medical device, and scientific instruments.
With Dimmable LED
This is equipped with high-light intensity LED lamps.
The temperature and humidity are controlled by a 3-position control system, contributing to energy savings. Optimal program control can be achieved based on the specific requirements of the cultured plants. All types are capable of continuous dimming.
LED Spectrum
Model | White LED | LH-411PFQST-S | LPH-411PFQST-S | LH-411PFQST-SP | LPH-411PFQST-SP | LPH-411PFQST-SPC |
---|---|---|---|---|---|---|
Yellow LED | LH-411PFQDT-S | LPH-411PFQDT-S | LH-411PFQDT-SP | LPH-411PFQDT-SP | LPH-411PFQDT-SPC | |
Pink LED | LH-411PFQPT-S | LPH-411PFQPT-S | LH-411PFQPT-SP | LPH-411PFQPT-SP | LPH-411PFQPT-SPC | |
Outer dimensions (mm) | W880 × D806 × H1,935 | |||||
Inner dimensions (mm) | W670 × D565 × H1,100 |
|||||
Control system |
Changeover between constant operation |
Real-time based control (24 hours)/Cumulated hours-based control(Up to 999 hours and 59 minutes/step) 24steps/5patterns with a pattern link function |
||||
Repeat | ― | 1 to 99 times or unlimited | ||||
Control methods | Three-position control method: Refrigerator and humidifier, ON/OFF control; heater, proportional control | |||||
Temperature | +5℃ to 50℃ ± 1℃ ( +15℃ ~ when all lights on) | |||||
Humidity | ― |
50-90%RH ± 10%RH |
― | 50-90%RH ± 10%RH (15℃ to 45℃) |
||
CO2 concentration | ― |
Ambient to 3,000 ppm |
||||
Illuminance1 | White: 0 to 58,500 lx, Yellow: 0 to 76,000 lx, Pink: 0 to 18,500 lx They can be continuously dimmed from 20% to 100% or 0% (all lights off) |
|||||
Photon flux density1 | White:0 to 900 μmolm-2s-1, Yellow: 0 to 930 μmolm-2s-1, Pink: 0 to 640 μmolm-2s-1 | |||||
Light source |
Illumination |
White: Straight tube sun-like LED lamp (SUNRAY LIGHT) |
||||
Number of lamps | Left, right, rear, and door: 4 lamps (40W) x 4 sides, Celling: 1 lamp (600W) x 1 side | |||||
Refrigerator | 300W | |||||
Heater | 500W | |||||
Humidifier | ― | 80W | ― | 80W | ||
Shelf | 6 shelves (adjustable) | |||||
Operation current |
White: 17A |
White: 20.5A |
White: 17A |
White: 20.5A |
||
Power requirement |
Single phase 100V |
Single phase 100V |
Single phase 100V |
Single phase 100V |
||
Weight (kg) | About 295 | About 305 | About 295 | About About 305 |
1 Illuminance and photon flux are measured on the center shelf inside the cabinet.
* Temperature sensor: platinum resistance temperature sensor, humidity sensor: capacitance measurement sensor, humidifier: ultrasonic humidifier.
* Power requirement does not include the power-outlet installed inside the chamber.
* These chambers are designed to operate in ambient temperatures up to 30℃. The specified performance may not be achieved depending on actual usage conditions.
* If lighting, temperature setting should be 15℃ or higher.
* The specified performance may not be achieved depending on the combination of temperature and humidity.
* Depending on the sample, drainage installation may be required.