Totan Booth VCM-1500N2


model VCM-1500N2

FEATURES
Internal pressure is maintained negative to prevent outflow of aerosol generated during
phlegm collection.
Vertical laminar airflow system is installed.
Cleanliness:ISO Class5 can be maintained inside the booth.
Secures 250 ventilations per hour to quickly eliminate contaminated aerosol in the booth.
Air is supplied evenly from the whole ceiling to prevent engulfing or adverse current of @ aerosol during phlegm collection.
HEPA filters both for supplying and exhausting air are installed. The one for air exhaust contains sterilizing enzyme to prevent outflow of contaminated aerosol.
One fluorescent light(30W x 1) and one sterillizing lamp(15W x 1) are equipped.
The fan can be set either automatic or manual operation. Under the automatic operation, a photoelectronic sensor detects the patient in front of the door to start the fan for negative pressure control, and sterilization lamp is irradiated to sterilize the booth interior when the room is not used.
@
SPECIFICATIONS
Model VCM-1500N2
Outer Dimension(mm) W1,500~D900~H2,175
Internal Dimension(mm) W1,150~D794~H1,823
Filter Air supply HEPA filter(dust collection efficiency:PAO particles of 0.3m
99.99% or higher
Pre-filter(Nonwoven nylon)
Air exhaust HEPA filter with sterillizing enzyme(dust collection efficiency:
PAO particles of 0.3m,99.99% or highj
Pre-filter(Antimicrobial PP honeycomb net)
Ventilation frequency 250 times/h
Fluorescent light 30W~P
Sterillizing lamp 15W~1
Internal pressure Negative
Power supply Single phase 100u@15A